Industrial Kelvin Double Bridge

Model : 55X-IK

 

DESCRIPTION-

“RAI” Industrial Kelvin Double Bridge Model 55X-IK is an important instrument for the measurement of low resistance values ranging from 0.2 Micro-ohms to 11 ohms. This kelvin double bridge is designed for accurate low-resistance measurement and is suitable for industrial, laboratory, calibration, and engineering applications.

The Industrial Kelvin Bridge employs 10 standard resistors made from well-aged manganin sheets having excellent stability and reproducibility. Manganin has a very low temperature coefficient, which is a critical parameter in the manufacturing of resistance standards. Four standard decade dials, ratio selector, and plug keys facilitate accurate null detection.

The instrument operates on the principle of a kelvin bridge, using a four-terminal measurement method to minimize the effects of lead and contact resistance. The arrangement of the bridge circuit, resistance arms, ratio settings, and measurement terminals can be understood through a kelvin double bridge diagram, which illustrates the low-resistance measurement principle and connection method.

SPECIFICATIONS

Parameter

Specification

Model

55X-IK

Instrument

Industrial Kelvin Double Bridge

Range of Measurement

0.2 Micro-ohm to 11 ohm

Current Source

10 Amp DC

Variable Standard Resistance

10 steps of 0.01 ohm

Multiplying Ratio

0.01, 0.1, 1, 10 and 100

Slide Wire

0.001 ohm divided into 500 equal divisions

Switches

Hard silver plated, giving constant values up to 10,000 operations

Measurement Method

Four-terminal low-resistance measurement

Resistance Material

Well-aged Manganin

Application

Low Resistance Measurement and Calibration

Accuracy

Resistance Range

Accuracy

Below 0.0001 ohm

±0.2% ±1 slide wire division

0.0001 ohm to 0.01 ohm

±0.1%

Above 0.01 ohm

±0.05%

Range Factor, Total Range & Resolution

Range Factor Setting

Total Range

Resolution

0.01

0.0011 ohm

0.2 micro-ohm

0.1

0.011 ohm

2 micro-ohm

1.0

0.11 ohm

20 micro-ohm

10

1.1 ohm

0.2 milli-ohm

100

11.0 ohm

2.0 milli-ohm

FEATURES

  • Precision measurement range from 0.2 micro-ohm to 11 ohm.

  • Four-terminal method for accurate low-resistance measurement.

  • Manganin wire having an exceptionally low temperature coefficient.

  • Remarkably long-lasting resistance values with excellent stability.

  • Highly reproducible measurements.

  • Perfectly aged Manganin coils/strips employed in the kelvin double bridge.

  • Click-type, hard silver-plated, self-cleaning switches.

  • Excellent elastic properties of Manganin.

  • Solid brass blocks for positive plug contacts.

  • Excellent calibration stability.

  • Designed for accurate low-resistance measurement using the kelvin bridge principle.

  • Suitable for applications requiring precise and stable low-resistance measurement.

ACCESSORIES FOR INDUSTRIAL KELVIN DOUBLE BRIDGE

The following accessories are required with the Industrial Kelvin Double Bridge as support accessories to complete the measurement system. The Bridge becomes a complete measurement system in conjunction with the following accessories:

  • Conductivity Attachment

  • High Amperage DC Source (0 to 50 Amp. suitable for bridge)

  • Spot Reflecting Galvanometer

  • Analogue Galvanometer with selector switch for protection of Spot Reflecting Galvanometer

Frequently Asked Questions

1. What is a Kelvin Double Bridge?

A Kelvin Double Bridge is a precision electrical instrument used for measuring very low resistance values while minimizing the effects of lead and contact resistance. The 55X-IK uses a four-terminal measurement method for accurate low-resistance measurement.

2. What is the measurement range of the 55X-IK?

The Industrial Kelvin Double Bridge Model 55X-IK measures resistance from 0.2 micro-ohms to 11 ohms, making it suitable for precise low-resistance measurement and calibration applications.

3. What is the accuracy of the Industrial Kelvin Double Bridge?

The 55X-IK provides accuracy of ±0.2% ±1 slide wire division below 0.0001 ohm, ±0.1% from 0.0001 ohm to 0.01 ohm, and ±0.05% above 0.01 ohm.

4. How does a Kelvin Bridge measure low resistance accurately?

A kelvin bridge uses a four-terminal measurement arrangement to reduce errors caused by lead and contact resistance. The ratio arms, standard resistance, current source, and null detection arrangement work together to provide accurate low-resistance measurements.

5. Where is the 55X-IK Industrial Kelvin Double Bridge used?

The 55X-IK is suitable for industrial, laboratory, calibration, and engineering applications requiring accurate low-resistance measurement. A kelvin double bridge diagram can also be used to understand its circuit arrangement and measurement principle.